KEYENCE DV-90 User Manual
Page 192
8-14
E DV-90-IM
8
Ser
ia
l Comm
unicati
on
8-3 Command Communication Details
Settings relevant to PORT1/PORT2
Function
Sending command
Response
Notes
PORT Unit
Change
WRD0mn
OK
m = 0: PORT1
1: PORT2
n = 0: none
1: PC (PORT2 only)
2: PLC link (PORT2 only)
3: Code Reader
Check
RDD0m
n
Baud rate
Change
WRD1mn
OK
m = 0: PORT1
1: PORT2
n = 0: 600 bits/s
1: 1200 bits/s
2: 2400 bits/s
3: 4800 bits/s
4: 9600 bits/s
5: 19200 bits/s
6: 31250 bits/s
7: 38400 bits/s
8: 57600 bits/s
9: 115200 bits/s
Check
RDD1m
n
Data length
Change
WRD2mn
OK
m = 0: PORT1
1: PORT2
n = 0: 7 bits
1: 8 bits
Check
RDD2m
n
Parity
Change
WRD3mn
OK
m = 0: PORT1
1: PORT2
n = 0: None (n)
1: Even (e)
2: Odd (o)
Check
RDD3m
n
Stop bit length
Change
WRD4mn
OK
m = 0: PORT1
1: PORT2
n = 0: 1 bit
1:2 bits
Check
RDD4m
n
RTS/CTS Protocol
Change
WRD5mn
OK
m = 1: PORT2 (Fixed)
n = 0: OFF
1: ON
Check
RDD5m
n
Input Limit
Change
WRD6mn
OK
m = 0: PORT1
1: PORT2
n = 0: OFF
1: Forward
2: Reverse
Check
RDD6m
n
Limit Start
Change
WRD7mnnnn
OK
m = 0: PORT1
1: PORT2
nnnn = 0001 through 1000: limit start
Check
RDD7m
nnnn
Limit Digits
Change
WRD8mnnn
OK
m = 0: PORT1
1: PORT2
nnn = 001 through 500: limit digits
Check
RDD8m
nnn