Links – Metrohm IC Net 2.3 User Manual
Page 99
6.5 761 Compact IC
IC Net 2.3
89
761 unit version
Selection of instrument version:
761.0010
= 761 Compact IC without sup-
pressor
761.0020
= 761 Compact IC with suppres-
sor
Suppressor
Suppressor
module:
Autostep with
Automatic switching to the next position
each time the injection valve is switched ei-
ther to the
Fill
or to the
Inject
position.
Use peristaltic pump
If this option is disabled, the peristaltic
pump is not switched on automatically at
the start of a determination or with
Startup
hardware
.
Polarity
Selection of the output signal polarity:
+
positive polarity (for anions)
–
negative polarity (for cations)
Measure channel
Display of the data source selected in the
Data source
window (see section 6.26.2).
Links
The
Links
tab of the
761 Compact IC
is used for COM port selection
and settings (details see section 5.2.4 Links).