Pulse mode – Metrohm IC Net 2.3 User Manual
Page 117
![background image](/manuals/586387/117/background.png)
6.7 871 Advanced Bioscan
IC Net 2.3
107
Startup values
E
Electric potential applied to the amperomet-
ric detector cell in V
Range:
-2 ... +2 V
.
Range
Selection of measuring range:
pA
10 ... 500 pA
.
nA
1 ... 500 nA
.
µA
1 ... 200 µA
.
Filter
Electronic damping of the analog output
signal.
Selection:
off, 0.2, 0.5, 1, 2, 5 s
Offset
Offset of the zero point for the analog out-
put signal.
Selection:
-50 ... +50 %
(in steps of 10 %)
Data rate
Setting of the data acquisition rate.
Selection:
-1, 2, 5, 10 Hz
Polarity
Set sign:
+
or
-
Actual
E
Display of current cell potential in V
Range
Display of current measuring range.
Filter
Display of current damping constant.
Offset
Display of current offset.
Data rate
Display of current data rate.
Polarity
Display of current polarity.
Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (
*.smt
) as long as
the file is not saved.
Pulse mode
This subtab of the
Method parameters
tab in the
871 Advanced Bio-
scan
window contains the method parameters for the Pulse
mode. These values are sent and applied to the 871 Advanced
Bioscan by
Start determination
and
Startup hardware
, or if the val-
ues are sent manually with
.
Startup values
E1, E2, E3
Definition of three different working poten-
tials, which are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the
Method
op-
tion is enabled above the graphic window.
Range:
-2 ... +2 V
.
t1, t2, t3
Time duration to apply the working poten-
tial.
Range:
0.1 ... 2 s
(t1)
, 0.01 … 2 s
(t2,t3)