6.7 871 Advanced Bioscan

IC Net 2.3

107

Startup values

E

Electric potential applied to the amperomet-
ric detector cell in V
Range:

-2 ... +2 V

.

Range

Selection of measuring range:

pA

10 ... 500 pA

.

nA

1 ... 500 nA

.

µA

1 ... 200 µA

.

Filter

Electronic damping of the analog output
signal.
Selection:

off, 0.2, 0.5, 1, 2, 5 s

Offset

Offset of the zero point for the analog out-
put signal.
Selection:

-50 ... +50 %

(in steps of 10 %)

Data rate

Setting of the data acquisition rate.
Selection:

-1, 2, 5, 10 Hz

Polarity

Set sign:

+

or

-

Actual

E

Display of current cell potential in V

Range

Display of current measuring range.

Filter

Display of current damping constant.

Offset

Display of current offset.

Data rate

Display of current data rate.

Polarity

Display of current polarity.

Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (

*.smt

) as long as

the file is not saved.

Pulse mode

This subtab of the

Method parameters

tab in the

871 Advanced Bio-

scan

window contains the method parameters for the Pulse

mode. These values are sent and applied to the 871 Advanced
Bioscan by

Start determination

and

Startup hardware

, or if the val-

ues are sent manually with

.

Startup values

E1, E2, E3

Definition of three different working poten-
tials, which are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the

Method

op-

tion is enabled above the graphic window.
Range:

-2 ... +2 V

.

t1, t2, t3

Time duration to apply the working poten-
tial.
Range:

0.1 ... 2 s

(t1)

, 0.01 … 2 s

(t2,t3)