List of program instructions, Links – Metrohm IC Net 2.3 User Manual
Page 159
6.14 753 Suppressor Module
IC Net 2.3
149
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see be-
low).
ENABLED
Enable program start (a disabled program
is not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
page:
Instruction Parameter
entry Meaning
Program RESET,
END
Set program flag. The
END
flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration
. Additional
steps after this flag are not al-
lowed. The
RESET
flag is used to
reset the parameters to the sys-
tem startup values.
Pump on,
off
Switch on or off the pump drive.
Step
Rotate suppressor to the next
position.
Links
The
Links
tab of the
753 Suppressor Module
is used for Event line
selection.