6.14 753 Suppressor Module

IC Net 2.3

149

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see be-
low).

ENABLED

Enable program start (a disabled program
is not started).

Add new program instruction.

Delete selected program instruction.

Test the time program (error messages are
displayed if program is wrong).


List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

page:

Instruction Parameter

entry Meaning

Program RESET,

END

Set program flag. The

END

flag

can be used to end a program,

especially if the program time

should be longer than the chro-

matogram

duration

. Additional

steps after this flag are not al-

lowed. The

RESET

flag is used to

reset the parameters to the sys-

tem startup values.

Pump on,

off

Switch on or off the pump drive.

Step

Rotate suppressor to the next

position.

Links

The

Links

tab of the

753 Suppressor Module

is used for Event line

selection.