Time program – Metrohm IC Net 2.3 User Manual
Page 251

6.27 861 Advanced Compact IC
IC Net 2.3
241
Peristaltic
Peristaltic
pump.
(inactive with version 2.861.0010)
Start peristaltic pump.
Stop peristaltic pump.
CO
2
Suppressor
853
CO
2
Suppressor
(only displayed with version 2.861.0040)
Start 853 CO
2
Suppressor.
Stop 853 CO
2
Suppressor.
System startup values
The system startup values are sent and ap-
plied to the 861 Advanced Compact IC 
each time the system is connected, a de-
termination is started, or the values are sent 
manually with 
.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump. 
Entry range: 
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit 
for high-pressure pump. This limit is con-
trolled even without connection to the PC. 
Entry range: 
0.0
...
35.0 MPa
If the
Flow
is higher than 1.5 mL/min, the
defined
Pressure max
must not exceed
25.0 MPa (250 bar).
Pressure min, MPa
Startup value for minimum pressure limit for 
high-pressure pump. This limit is controlled 
even without connection to the PC. 
Entry range: 
0.0
...
35.0
MPa
Full scale,
µ
S/cm
Startup value for full scale range. 
Selection: 
50
,
250
,
1000, 5000
µ
S/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1
Time program
On the
Program
tab of the
861 Advanced Compact IC
window a
user-defined time program for the remote output lines of the 861 
Advanced Compact IC can be entered. This program is started 
automatically as defined in the 
Start mode
window (see section
4.4.3) either at the moment the determination is started (
Start with
determination
) or at the moment the sample is injected (
Start with
inject
).