beautypg.com

Time program – Metrohm IC Net 2.3 User Manual

Page 251

background image

6.27 861 Advanced Compact IC

IC Net 2.3

241

Peristaltic

Peristaltic

pump.

(inactive with version 2.861.0010)

Start peristaltic pump.

Stop peristaltic pump.

CO

2

Suppressor

853

CO

2

Suppressor

(only displayed with version 2.861.0040)

Start 853 CO

2

Suppressor.

Stop 853 CO

2

Suppressor.

System startup values

The system startup values are sent and ap-

plied to the 861 Advanced Compact IC
each time the system is connected, a de-
termination is started, or the values are sent
manually with

.

Flow, mL/min

Startup value for flow rate of the high-
pressure pump.
Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit
for high-pressure pump. This limit is con-
trolled even without connection to the PC.
Entry range:

0.0

...

35.0 MPa

If the

Flow

is higher than 1.5 mL/min, the

defined

Pressure max

must not exceed

25.0 MPa (250 bar).

Pressure min, MPa

Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:

0.0

...

35.0

MPa

Full scale,

µ

S/cm

Startup value for full scale range.
Selection:

50

,

250

,

1000, 5000

µ

S/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1

Time program

On the

Program

tab of the

861 Advanced Compact IC

window a

user-defined time program for the remote output lines of the 861
Advanced Compact IC can be entered. This program is started
automatically as defined in the

Start mode

window (see section

4.4.3) either at the moment the determination is started (

Start with

determination

) or at the moment the sample is injected (

Start with

inject

).