Controls – Metrohm IC Net 2.3 User Manual
Page 260

6 Devices
IC Net 2.3
250
Controls
On the subtab
Controls
injection valve, peristaltic pump, high-
pressure pump, column heating and remote lines can be manu-
ally controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:
Valve
Injection
valve
Switch injection valve to "Inject" position.
Switch injection valve to "Fill" position.
Peristaltic pump
Peristaltic
pump
Start peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
Stop peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
IC pump
High pressure pump
Start high pressure pump.
Stop high pressure pump.
Flow, mL/min
Set value for flow rate of IC pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Value for maximum switch-off pressure for
high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under
IC NET / Global
preferences
).
Entry range:
0.0
...
35.0 MPa
If the
Flow
is higher than 1.5 mL/min, the
defined
Pressure max
must not exceed