6 Devices

IC Net 2.3

250

Controls

On the subtab

Controls

injection valve, peristaltic pump, high-

pressure pump, column heating and remote lines can be manu-
ally controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:

Valve

Injection

valve

Switch injection valve to "Inject" position.

Switch injection valve to "Fill" position.

Peristaltic pump

Peristaltic

pump

Start peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).

Stop peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).

IC pump

High pressure pump

Start high pressure pump.

Stop high pressure pump.

Flow, mL/min

Set value for flow rate of IC pump.
Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Value for maximum switch-off pressure for

high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under

IC NET / Global

preferences

).

Entry range:

0.0

...

35.0 MPa

If the

Flow

is higher than 1.5 mL/min, the

defined

Pressure max

must not exceed