Outputs – Metrohm IC Net 2.3 User Manual
Page 257
6.27 861 Advanced Compact IC
IC Net 2.3
247
For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.
Possibility for changing the break times for
injection valve
Valve
and suppressor mod-
ule
Suppressor
.
Outputs
The
Outputs
tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Suppressor step
Switching of the suppressor module to the
next position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 861 Ad-
vanced Compact IC (high-pressure pump,
injection valve, or suppressor not working
correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
Remote line
Set remote output lines 1
…
8.
Selection:
0
(line off, inactive, open)