6.27 861 Advanced Compact IC

IC Net 2.3

247

For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.

Possibility for changing the break times for
injection valve

Valve

and suppressor mod-

ule

Suppressor

.

Outputs

The

Outputs

tab of the hardware settings window defines remote

output signals to be set automatically if specific events occur.

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "IN-
JECT" position.

Fill

Switching of the injection valve to the "FILL"
position.

Suppressor step

Switching of the suppressor module to the
next position.

Pump stop (P max/min)

Pump stopped because pressure limits are
exceeded.

Hardware error

Hardware error detected at the 861 Ad-
vanced Compact IC (high-pressure pump,
injection valve, or suppressor not working
correctly).

Alarm leak detector

Leak detector has detected solvent in the

instruments interior.

Remote line

Set remote output lines 1

8.

Selection:

0

(line off, inactive, open)