6 Devices

IC Net 2.3

242

The

Program

tab contains the two following subpages:

Program

Main time program with all program
steps.

Remote configuration

Possibility for creation of user-defined
remote commands.

Program

On the

Program

subpage, program steps including time, program

instruction and parameter can be entered.

First column

Time at which program instruction is app-
lied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the

Remote configura-

tion

tab.

Third column

Parameter for program instruction (see be-
low).

ENABLED

Enable program start (a disabled program
is not started).

Add new program instruction.

Delete selected program instruction.

Test the time program (error messages are
displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

subpage: