Mix method / user program – Metrohm IC Net 2.3 User Manual
Page 225

6.25 Metrohm Spark Triathlon Autosampler
IC Net 2.3
215
Apply values selected under
Method values
to the
End time
, the set values are displayed
under
Current
.
For detailed information about the I/O connections of the Metrom 
SparkTriathlon Autosampler see chapter 6 of the Spark Triath-
lon user manual. 
Mix method / User Program
The
Methods programming/Triathlon methods/Mix
and the
Methods
programming/User program
tab are structured the same way. They
both allow configuring of liquid handling methods. The difference 
is the set of actions that is available. The 
Mix method
is restricted
to liquid handling before the injection, e.g. mixing or dilution, while 
in the 
User program
the whole set of actions of the
Triathlon Auto-
sampler
is available.