5 762 event output lines, Time program – Metrohm IC Net 2.3 User Manual
Page 56
5 Interfaces
IC Net 2.3
46
Put on desktop
Possibility for setting COM port parameters
(details see on-line help).
Change
Possibility for changing the COM port. The
following window is opened, where the
COM port can be changed by clicking on
the desired entry.
5.2.5
762 Event output lines
762 System [#] icon / Open
The
762 System [#]
window with programs and settings for the
seven event output lines is opened by selecting this menu option
with the right mouse button or by double-clicking the
762 icon
in
the
SYSTEM
window. It consists of the tabs
Program
and
Events
setup
.
Time program
On the
Program
tab of the
762 System [#]
window a user-defined
time program for the event output lines of the 762 IC Interface can
be entered. This program is started automatically as defined in
the
Start mode
window (see section 4.4.3) either at the moment
the determination is started (
Start with determination
) or at the
moment the sample is injected (
Start with inject
).