Metrohm IC Net 2.3 User Manual
Page 350

8 Chromatograms 
 
IC Net 2.3
340
[Item]
TableStart =
A6, Specifies the cell in the
data
worksheet of the
ExcelReport.xls
where the header of the first col-
umn is located. Defines the position of the data 
cells of the 
data
worksheet.
For correct display of the data in the
ExcelReport.xls the for-
matting of the cells has to match the containing data type. Re-
member to adjust the formatting when changing the value of 
TableStart, important especially for correct display of 
DateTime data. 
NoItems =
n
, number of items that are to be displayed in the
data
worksheet.
Item1 =
Keyword
. The
Itemx
entry specifies the columns of
the
data
worksheet. The numbering
1 – n
after
Item
specifies the order of the columns, the
Key-
word
after
=
specifies what kind of data is ex-
tracted from the «IC Net» report and filled in the 
columns. 
Itemn =
Keyword:
No
Peak number.
Retention
Retention time of the component (in 
minutes) 
Width/2
Width of the peak at half height (in 
minutes). 
Height
Height of the peak (in µS/cm).
Area
Area of the peak.
K’
Capacity factor, ratio of corrected re-
tention time to the void time. 
Resolution
Resolution for two neighboring peaks
R(i,i+1)
.
TP
Effectivity for the peak in number of 
theoretical plates. 
TP/m
Effectivity for the peak in number of 
theoretical plates per meter. 
HETP/m
Height of theoretical plate divided into 
particle size. 
Asym.
Asymmetry of the peak measured at 
1/10 of its height. 
RF
Response factor, coefficient K1 of the 
calibration curve. 
Conc.
Absolute
concentration.
. 
. 
. 
. 
. 
. 
. 