beautypg.com

Common – Metrohm IC Net 2.3 User Manual

Page 114

background image

6 Devices

IC Net 2.3

104

Scan

The

Scan

mode allows potential/current

curves to be recorded.

Cell

Manual control of amperometric detector
cell:

Switch on detector cell.

Switch off detector cell.

Oven

Manual control of column oven (heater):

Switch on oven.

Switch off oven.

Temperature

Temperature of column oven.
Range:

15 ... 45 °C

Control

Manual setting of 4 control lines.

Set control lines to selected states.

Aux1

,

Aux2

,

Cont1

,

Cont2

Setting of 4 control lines.
Selection:

0

,

1

,

*

Valve

Manual switching of the injection valve at
the 812 Valve Unit:

Switch injection valve to "Fill" position (con-
trol line

Cont1

is set to 0).

Switch injection valve to "Inject" position
(control line

Cont1

is set to 1).

Autozero

Manual setting of the analog output signal
to zero.

Turn on autozero function.

Turn off the autozero function.

Actual

Display of actual values.

Mode

Display of current detector mode.

Cell

Display of current cell state.

Oven

Display of current ovent state.

Control

Display of current control line settings.

Valve

Display of current valve position.

Autozero

Display of current autozero state.

Common

This tab of the

871 Advanced Bioscan

window contains the method

startup values for the common parameters of the instrument.
These values are sent and applied to the 871 Advanced Bioscan
by

Start determination

and

Startup hardware

, or if the values are

sent manually with

.