6 concentration table – Metrohm IC Net 2.3 User Manual
Page 305

7.5 Calibration and quantification
IC Net 2.3
295
Worst case...
Information on the component with the 
worst (largest) deviation of actual and ex-
pected retention time. Deviation is given as 
a part of component’s identification win-
dow. 
Average relative deviation
Relative deviation averaged for all compo-
nents. 
7.5.6 Concentration
table
IC NET / Method / Calibration / Concentrations
Selecting this menu item or clicking on
in the
component table opens the
Concentrations
window with the con-
centration table containing concentrations of all components for 
all 
Calibration levels
. Each calibration level corresponds to a sam-
ple used for calibration and to a point on the calibration curve.
Concentration units
User defined units of concentration which 
will appear in the report. Nevertheless, entry 
of new concentration units does not cause 
recalculation of concentrations.
Data type
Choice of the data type that will be shown 
in the concentration table: 
concentrations
(default setting), peak
heights
, peak
areas
,
volumes
.
Number
Number of the component (read-only).
Name
Name of the component taken from the 
component table (read-only). 
This run
Contains concentrations (or other chosen 
values) obtained in the current run (read-
only, except for universal component). 