Time program – Metrohm IC Net 2.3 User Manual
Page 126
6 Devices
IC Net 2.3
116
Entry field
Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection:
fill, inject, off
* only available for
733 IC Separation Center
connected to 732
Time program
The
Program
tab is only available for a
733.0XX0 Separation Center
connected to a 762 IC Interface. The
733 IC Separation Center
that is connected to a 732 IC Detector is controlled by the time
program of the detector.
On the
Program
tab of the
733.0XX0 Separation Center
window a
user-defined time program can be entered. This program is
started automatically as defined in the
Start mode
window (see
section 4.4.3) either at the moment the determination is started
(
Start with determination
) or at the moment the sample is injected
(
Start with inject
).
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see be-
low).
ENABLED
Enable program start (a disabled program
is not started).