Links – Metrohm IC Net 2.3 User Manual
Page 229
![background image](/manuals/586387/229/background.png)
6.25 Metrohm Spark Triathlon Autosampler
IC Net 2.3
219
Position number of the transport vial, is
successively increased by one as a sam-
ple queue is processed.
Transport Vial = specified sample Vial +
Combo box to specify the position of the
transport vial relative to the actual sample
vial position. If activated,
First Transport Vial
is disabled.
Entry range:
max.
-95…+95
, depends on
type of tray segment.
Reagent A…D
Specifies the reagent vials, up to four re-
agents can be defined.
Use
Enable/disable the usage of the reagent vial
in a
User Program/Mix method
.
Segment type
Define the type of tray segment that con-
tains the reagent vial.
A
Standard tray segment,
B
LSV tray segment,
C
Super LSV tray segment,
D
Micro vial tray segment,
Position
Define the position number of the vial.
Entry range:
1…max. 96
, depends on type of
tray segment.
Apply values selected, the set values are
displayed under
Current
.
For detailed information about the Triathlon Autosampler see
Spark Triathlon user manual.
Links
The
Links
tab of the
Triathlon Autosampler
is used for COM port
selection and settings (details see section 5.2.4 Links).