6.28 844 UV/VIS Compact IC

IC Net 2.3

255

Delete selected program instruction.

Test the time program (error messages are
displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

subpage:

Remote

0, 1, *, p

Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter

1

,

0

,

p

or

*

. For entry of the

other values, move the cursor in front
of the value to be changed and enter

1

,

0

,

p

or

*

.

Valve Inject,

Fill

Switch injection valve to "INJECT"
or "FILL" position.

ICPump on,

off

Switch on or off the high-pressure
pump
.

Flow

0.2 ... 2.5
mL/min

Set flow rate of the high-pressure
pump to the desired value.

Pmax

0.0 ... 35.0 MPa

Set maximum pressure limit for the
high-pressure pump to the desired
value.

Pmin

0.0 ... 35.0 MPa

Set minimum pressure limit for the
high-pressure pump to the desired
value.

Peristaltic on,

off

Switch on or off the peristaltic
pump
.

Wavelength_1..3

190 ... 900 nm

Set wavelength of the corresponding
measuring channel on the desired
value.

Bunching_1..3

1 ... 101 nm

Set bunching of the corresponding
measuring channel on the desired
value.

UV_Lamp on,

off

Switch on or off the UV lamp.

VIS_Lamp on,

off

Switch on or off the VIS lamp.

Zero

Record a new zero spectrum.