List of program instructions – Metrohm IC Net 2.3 User Manual
Page 265

6.28 844 UV/VIS Compact IC
IC Net 2.3
255
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Remote
0, 1, *, p
Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter
1
,
0
,
p
or
*
. For entry of the
other values, move the cursor in front
of the value to be changed and enter
1
,
0
,
p
or
*
.
Valve Inject,
Fill
Switch injection valve to "INJECT"
or "FILL" position.
ICPump on,
off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5
mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 35.0 MPa
Set maximum pressure limit for the
high-pressure pump to the desired
value.
Pmin
0.0 ... 35.0 MPa
Set minimum pressure limit for the
high-pressure pump to the desired
value.
Peristaltic on,
off
Switch on or off the peristaltic
pump.
Wavelength_1..3
190 ... 900 nm
Set wavelength of the corresponding
measuring channel on the desired
value.
Bunching_1..3
1 ... 101 nm
Set bunching of the corresponding
measuring channel on the desired
value.
UV_Lamp on,
off
Switch on or off the UV lamp.
VIS_Lamp on,
off
Switch on or off the VIS lamp.
Zero
Record a new zero spectrum.