Metrohm IC Net 2.3 User Manual

Page 147

background image

6.12 Metrohm 709 Solvent Delivery Unit

IC Net 2.3

137

Program table

Number

Row number in the program table (read-
only).

Time

Time at which the flow and/or gradient is
changed.
Entry range:

0.0 ... 99999 min

Flow

Overall flow of pump control system in
mL/min.
Entry range:

0.01 ... 20.00 mL/min

(depending

on number of pumps)

%B

Individual flow rate for pump B in percent of
total flow.
Entry range:

0 ... 100 %

%C

Individual flow rate for pump C in percent of
total flow.
Entry range:

0 ... 100 %

%D

Individual flow rate for pump D in percent of
total flow.
Entry range:

0 ... 100 %

Enabled

Enable time program start (a disabled pro-
gram is not started).

Show flow column

Show

Flow

column in the program table.

Add a new line to the time program.

Delete the selected line from the time pro-
gram.

Start all pump drives.