13 752 pump unit, 1 752 pump unit features, 2 752 pump unit icon – Metrohm IC Net 2.3 User Manual
Page 154: 3 752 pump unit window
![background image](/manuals/586387/154/background.png)
6 Devices
IC Net 2.3
144
6.13
752 Pump Unit
6.13.1
752 Pump Unit features
The Metrohm 752 Pump Unit is a 2-channel peristaltic pump
which is normally used for operating the suppressor module in
the 733 IC Separation Center by supplying regeneration and rins-
ing solutions.
The 752 Pump Unit, which is normally operated manually using
the keypad of the instrument or automatically via 732 IC Detector
via remote interface, can be fully operated using the «IC Net»
program. The 752 Pump Unit is normally connected to one of the
event line connections of the 762 IC Interface using the optional
6.2128.180 cable. Examples for interconnections with the 752
Pump Unit can be found in the 762 Instructions for Use.
In order to operate the 752 Pump Unit with «IC Net», the external
control must be switched on using the [ Remote ] key.
For detailed information about the 752 Pump Unit see 752 In-
structions for Use.
6.13.2
752 Pump Unit icon
The
752 icon
is available in the
SYSTEM
window if a
752 Pump Unit
has been installed with the
New system wizard
or by using the
Setup/New devices/Install new device
option of the
SYSTEM
window
(see section 6.1.2).
If the system is connected and the
752 icon
is clicked with the
right mouse button, the following menu appears:
Open
Open
the
752 Pump Unit
window for parameter settings
(this window can also be opened by double-clicking
the icon).
Unlink
Delete
the
752 icon
from the system.
6.13.3
752 Pump Unit window
752 icon / Open
The
752 Pump Unit
window for parameter settings is opened by
selecting this menu option with the right mouse button or by dou-
ble-clicking the
752 icon
in the
SYSTEM
window. It consists of the
three tabs
Manual
,
Program
, and
Links
.