2 sample queue editor functions – Metrohm IC Net 2.3 User Manual
Page 357

9.3 Sample queue editor
IC Net 2.3
347
All system files of the sample queue must 
be stored in the same system subfolder. 
Ident
User defined identifier for chromatogram. 
Will be placed into appropriate passport 
field when starting the chromatogram. 
Vial
Autosampler vial position to take sample 
from. 
Volume
Injected volume of sample in
µ
L.
Dilution
Sample dilution prior to injection.
Amount
Sample amount. If this value is different for 
the calibration run (c) and the sample run 
(s), the component concentrations of the 
sample are calculated as follows: 
C
s
= C
c
•
Amount
s
/
Amount
c
Internal standard amount
Concentration of the internal standard 
component for relative concentration calcu-
lations. 
Level
Calibration level (see section 7.5) for the 
sample: Level 
0
stays for normal analysis,
levels
1
and greater for calibration runs.
Correct filling of this column enables to cal-
culate calibration coefficients automatically. 
Injections
Number of injections for the same vial posi-
tion (only 1 injection is permissible for 
Level
≥
1
).
Done
Indication whether the sample determina-
tion has been done or not: 
 
0
determination not yet done
1
…
n
determination or injection done
Sample Info 1
First sample description.
Sample Info 2
Second sample description.
 
 
 
 
9.3.2
Sample queue editor functions
The functions in the editor window for sample queues can be 
triggered with the corresponding menu items of the 
Edit
menu or
with the corresponding symbols in the symbol bar.
QUEUE EDITOR /
Edit / Undo
Undo the last modification of the sample queue table.
QUEUE EDITOR /
Edit / Cut row(s)