List of program instructions – Metrohm IC Net 2.3 User Manual
Page 83
6.3 732 IC Detector
IC Net 2.3
73
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter entry
Meaning
Remote
0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter
1
,
0
or
*
. For
entry of the other values, move
the cursor in front of the value to
be changed and enter
1
,
0
or
*
.
Flag reset,
end
Set program flag. The
end
flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration
. Additional
steps after this flag are not al-
lowed. The
reset
flag is used to
reset the parameters to the sys-
tem startup values.
Range
100, 200, 5000
µ
S/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05
µ
S/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on
Range
setting).
Zero on,
off
Switch on or off the auto-zero
function.
Polarity +,
–
Reset the polarity of the output
signal.
Mark
Trigger marking signal.
ValveA fill,
inject
Switch injection valve A to "
inject
"
or "
fill
" position.
ValveB fill,
inject
Switch injection valve B to "
inject
"
or "
fill
" position.
Suppressor
Switch suppressor module to the
next position.
PumpRS on,
off
Switch on or off the high-pressure
pump connected to the 732 IC
Detector.
Flow
0.01 ... 5 mL/min
Set flow rate of the high-pressure
pump to the desired value.