Time program – Metrohm IC Net 2.3 User Manual
Page 141
![background image](/manuals/586387/141/background.png)
6.11 818 IC Pump
IC Net 2.3
131
used. If the pump exceeds the preset limit
value during operation, it is switched off
immediately. At the same time, the
Pressure
field color in the 818 settings window and in
the
WATCH WINDOW
is changed to the
value set for
Out of range
.
The
set
minimum limit value should lie far
enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is
automatically switched off. At the same
time, the
Pressure
field color in the 818 set-
tings window and in the
WATCH WINDOW
is
changed to the value set for
Out of range
.
Send current parameters immediately to the
818 IC Pump (only available if pump is run-
ning). Parameters are not stored in the sys-
tem file (
*.smt
) as long as the file is not
saved.
Set current parameters and start 818 IC
Pump drive.
Stop 818 IC Pump drive.
Time program
The
Program
tab of the
818 IC Pump
window a user-defined time
program can be entered. This program is started automatically as
defined in the
Start mode
window (see section 4.4.3) either at the
moment the determination is started (
Start with determination
) or at
the moment the sample is injected (
Start with inject
).