Display – Metrohm IC Net 2.3 User Manual
Page 86

6 Devices
IC Net 2.3
76
Display
The
Display
tab of the
819 IC Detector
window is only available for a
connected system. It displays the current measurement values
and parameters settings of the 819 IC Detector.
Conductivity
Live display of measured conductivity
(auto-zero value) in
µ
S/cm. This field is
available for the
WATCH WINDOW
. The
watch window color is changed to the
Out
of range
color if the measured value is more
than 80% outside the
Full Scale
range
.
abs. Cond.
Live display of absolute conductivity value
in
µ
S/cm. This field is available for the
WATCH WINDOW
. The watch window color is
changed to the
Out of range
color if the
measured value is bigger or equal to 110%
of
Range.
Ion
Display of ion settings:
Anion
:
polarity = +, temperature coeff. = 2.5
Cation
:
polarity = –, temperature coeff. = 1.5
Range
Display of measuring range in
µ
S/cm.
Full
scale
Display of full scale range in
µ
S/cm.
Cell
const
. Display
of cell constant of the conductivity
cell in /cm.
Thermostat
Display of temperature within in the detec-
tor block.
Offset