Metrohm IC Net 2.3 User Manual
Page 150

6 Devices 
 
IC Net 2.3
140
The
Manual
tab of the
Pump Control Unit
window is only available
for a connected system.
Pressure
Live display of current pressure in MPa. 
This field is also available for the 
WATCH
WINDOW
.
Flow
Live display of current flow in mL/min. This 
field is also available for the 
WATCH WIN-
DOW
.
New flow
A new flow can be set by moving the scroll 
bar, by entering a value or by changing the 
value using the up and down arrows. 
Entry range: 
0.01 ... 20.00 mL/min
(depending
on number of pumps)
Min/Max pressure
The
minimum and maximum pressure limit
for the individual pump units can be set in 
steps of 0.1 MPa. 
Entry range: 
0.0 ... 50.0 MPa
.
%B
Individual flow rate for pump B in percent of 
total flow. 
Entry range: 
0 ... 100 %
%C
Individual flow rate for pump C in percent of 
total flow. 
Entry range: 
0 ... 100 %
%D
Individual flow rate for pump D in percent of 
total flow. 
Entry range: 
0 ... 100 %