6 Devices

IC Net 2.3

262

Operating hours

Operating hours of the lamps.

UV lamp, h

Display of the operating hours of the UV
lamp.

Set

operating hours of the UV lamp

to 0. Click

it after replacing the UV lamp.

VIS lamp, h

Display of the operating hours of the VIS
lamp.

Set

operating hours of the VIS lamp

to 0. Click

it after replacing the VIS lamp.

6.28.5 Spectrum

window

The

SPECTRUM WINDOW

can be opened with a right-hand mouse-click on

the 844 icon and then selecting the menu item

Spectrum window

, or via the

button on the subtab

Manual/Detector

.

A selection can be made from three different spectra, the absorption

spectrum, intensity spectrum and the zero spectrum. Please note that each

time you click on

Absorption

,

Intensity

or

Zero

the function of the

button will be triggered.

The 844 measures the current intensity spectrum and

sends it together with the last zero spectrum to have

been recorded to . The absorption spec-

trum is recalculated from these two spectra.

Absorption

The absorption (with the units mAU) is given on the y-

axis as a function of the wavelength (x-axis). The ab-

sorption values are calculated for all measured wave-

lengths and shown a s spectrum. The calculation for-

mula for the absorption value is shown here as an ex-