Metrohm IC Net 2.3 User Manual
Page 256

6 Devices
IC Net 2.3
246
The
Hardware
tab of the hardware settings window defines ge-
neral parameters which are set automatically at power on of the
861 Advanced Compact IC.
Remote lines after power on
The remote output lines 1
…
8 are set to this
values after power on or a manual stop with
Shutdown hardware
.
Selection:
0
,
1
Thermostat
Operating temperature of the conductivity
cell.
Selection:
25
,
30
,
35
,
40
,
45 °C
,
off
Thermostatting functions only if the ambient tempera-
ture is at least 5 °C lower than the operating tempera-
ture. It normally takes 30
…
60 min after power on until
a temperature stability of
±
0.01 °C is attained.
Flow correction
Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range:
0.9
...
1.09
The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:
Flow correction
=
Displayed flow rate
Measured flow rate
Cell constant
Cell constant of the conductivity cell for cor-
rect display of the absolute conductivity.
Enter the cell constant printed on the detec-
tor block into this field.
Range:
0.1
...
1000.0 /cm