6.28 844 UV/VIS Compact IC

IC Net 2.3

251

25.0 MPa (250 bar).

Pressure min, MPa

Value for minimum switch-off pressure for

high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under

IC NET / Global

preferences

).

Entry range:

0.0

...

35.0 MPa

Column oven

Column

oven

Turn column oven on (only enabled, if a
column oven is connected).

Turn column oven off (only enabled, if a
column oven is connected).

Temperature, °C

Set value for temperature in the column
oven.
Entry range:

0 ... 75 °C

Remote lines

Set values for remote line settings 1

8.

Auswahl:

0

,

1

Detector

On the subtab

Detector

the UV/VIS detector can be controlled

manually. The parameters for the measurement can also be set:

UV lamp

UV

lamp

Turn UV lamp on.

Turn UV lamp off.

VIS lamp

VIS

lamp

Turn VIS lamp on.