Timebase program – Metrohm IC Net 2.3 User Manual
Page 221

6.25 Metrohm Spark Triathlon Autosampler
IC Net 2.3
211
Method values
Combo box to select the parameters of the
Wash method
.
Method Parameter
Wash volume, µL
Entry range:
300…9999 µL
Wash between
Setting when the washing step takes place.
Vials
wash after a sample vial is processed
Injections
wash after every injection
Apply parameter values selected under
Method values
to the
Wash method
.
Timebase program
The
Methods programming/Triathlon methods/Timebase
subtab of
the Triathlon Autosampler window contains the configuration of
the pre-defined
Timebase program
.
The
Timbase program
is a programmable method to control exter-
nal devices connected to the Triathlon Autosampler via connector
P5 AUXILLIARIES or P3 TIMED OUTPUTS.
Auxiliaries
Subtab for the settings to control up to four
external devices
AUX-1…AUX-4
via connec-