4 report destination, 5 peak table – Metrohm IC Net 2.3 User Manual
Page 317
![background image](/manuals/586387/317/background.png)
7.6 Report output
IC Net 2.3
307
7.6.4 Report
destination
The following output devices can be selected simultaneously in
any combination as targets for report output:
Screen
Output report (without curves) to screen.
Printer
Output report to printer.
File
Output report to file (see section 7.6.7).
7.6.5 Peak
table
The following parameters can be selected for peak table output
with determination results:
Quantification method
Selection of the method to calculate con-
centrations of components:
Response normalization
The method normalizes the sum of re-
sponses for all peaks of the chromatogram
to the NORM factor entered in the
Total %
for normalization
field:
R
i
% = NORM
•
R
i
/
∑
R
i
.
All peaks are calculated and reported no
matter whether they have been recognized
or not.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
Area +
Area% or Height + Height%
,
Name
Normalized concentration
The method normalizes the sum of abso-
lute concentrations for all peaks of the
chromatogram to the NORM factor entered
in the
Total % for normalization
field:
C
i
% = NORM
•
W
i
(R
i
) /
∑
W
i
(R
i
)
As long as no universal component is de-
fined, only peaks with non-zero concentra-
tions are listed.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
Height
,
Area
,
Response factor
,
Concentration%
,
Name
Absolute concentration
This quantification method reports raw
quantity (absolute concentration) of the
component which is calculated directly
from the calibration formula:
Q
i
= W
i
(R
i
) / V'