List of program instructions – Metrohm IC Net 2.3 User Manual
Page 97

6.5 761 Compact IC
IC Net 2.3
87
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter
entry Meaning
Valve Inject,
Fill
Switch injection valve to "INJECT"
or "FILL" position.
FullScale
50, 250, 1000
µ
S/cm
Set full scale range to the se-
lected value.
ICPump on,
off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5 mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 25.0 MPa
Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin
0.0 ... 25.0 MPa
Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote
0, 1, *, p
Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter
1
,
0
,
p
or
*
. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter
1
,
0
,
p
or
*
.
Program END,
RESET
The
END
flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The
RESET
flag is used to reset the
parameters to the system startup
values.
Suppressor
Switch suppressor module to the
next position.
Peristaltic on,
off
Switch on or off the peristaltic
pump.
Measure_Off
Switch off measurement.