6.4 819 IC Detector
IC Net 2.3
81
Remote configuration
On the
Remote configuration
subtab user-defined remote com-
mands can be defined, which can be inserted into a time pro-
gram after being activated with
.
Name (1
st
column)
User-definable name of the remote com-
mand (e.g.
Start_766
).
Remote command (2
nd
column)
Setting the remote output lines 1
…
8.
0
(line off, inactive, open)
1
(line on, active, 0 V)
*
(leave line in current status)
For entry of the first value, enter
1
,
0
or
*
.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter
1
,
0
or
*
.
Add new remote command.
Delete selected remote command.
Activate the defined remote commands for
insertion into the time program.