Metrohm IC Net 2.3 User Manual
Page 115
6.7 871 Advanced Bioscan
IC Net 2.3
105
Mode
Selection of detector mode:
DC
In
DC
mode, a constant potential is applied
to the working electrode, the analytes are
oxidized or reduced in accordance with
their electrochemical properties.
Pulse
The
Pulse
mode works with three different
working potentials, which are applied cycli-
cally. This frees the electrode surface from
any adhering reaction products after each
sweep and an activated surface is pro-
duced for the next measurement.
Scan
The
Scan
mode allows potential/current
curves to be recorded.
Oven
Start value of column oven (heater):
Switch on oven.
Switch off oven.
Temperature
Temperature of column oven.
Range:
15 ... 45 °C
Control
Aux1
,
Aux2
,
Cont1
,
Cont2
Setting of 4 control lines.
Selection:
0
,
1
,
*