6.3 732 IC Detector

IC Net 2.3

69

6.3.3

732 IC Detector window

732 icon / Open

The

732 IC Detector

window for parameter settings is opened by

selecting this menu option with the right mouse button or by dou-
ble-clicking the

732 icon

in the

SYSTEM

window. It consists of the

four tabs

Display

,

Method parameter

,

Program

, and

Links

.

Display

The

Display

tab of the

732 IC Detector

window is only available for a

connected system. It displays the current measurement values
and parameters settings of the 732 IC Detector.

Conductivity

Live display of measured conductivity
(auto-zero value) in

µ

S/cm. This field is

available for the

WATCH WINDOW

. The

watch window color is changed to the

Out

of range

color if the measured value is more

than 80% outside the

Full Scale

range

.

abs. Cond.

Live display of absolute conductivity value
in

µ

S/cm. This field is available for the

WATCH WINDOW

. The watch window color is

changed to the

Out of range

color if the

measured value is bigger or equal to 110%
of

Range.

Ion

Display of ion settings:

Anion

:

polarity = +, temperature coeff. = 2.5

Cation

:

polarity = –, temperature coeff. = 1.5