4 hardware settings, Outputs, Alarm stops – Metrohm IC Net 2.3 User Manual
Page 268

6 Devices
IC Net 2.3
258
6.28.4 Hardware
settings
Selection of the
Hardware
menu item using the right mouse button
opens the hardware settings window consisting of the four tabs
Outputs
,
Alarm stops
,
System settings
and
Detector settings
.
Outputs
The
Outputs
tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 844 UV/VIS
Compact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
PowerOn values
Display of the power on startup values for
remote output lines set on the
Hardware
tab.
Pulse length
Length of a pulse in ms.
Alarm stops
The
Alarm stops
tab of the hardware settings window defines the
events for which the 844 UV/VIS Compact IC is stopped immedi-
ately. At an alarm stop, high-pressure pump and peristaltic pump