Metrohm IC Net 2.3 User Manual
Page 362

10 Batch reprocessing 
 
IC Net 2.3
352
switched on) by updating the concentration 
table. 
Default scheme
Default setting for recalibration reprocess-
ing. The two options 
Apply final calibration...
and
Forget calibration points...
are switched
on. A new calibration is performed with the 
calibration runs and the resulting new cali-
bration parameters (component table, con-
centration table and calibration curve) are 
applied to all sample runs. 
Apply final calibration to all reprocessed files
Apply the updated calibration to all calibra-
tion and sample runs. If this option is 
switched off, the calibration stored in the 
first chromatogram is used for all other 
chromatograms. 
Forget calibration points before reprocessing
Forget old calibration points of the calibra-
tion curve and perform a new calibration 
using all calibration runs of the batch re-
processing queue. In this case each cali-
bration chromatogram adds a new point to 
the calibration curve. If this option is 
switched off, the calibration curve stored in 
the first chromatogram remains active. 
Each further calibration chromatogram in 
the batch reprocessing queue adds a new 
point to this calibration curve. 
Update retention time on runs:
Update the retention times in the peak table 
after the run. Tool to analyse difficult sam-
ples. Compensates a drift of the retention 
time to enable correct peak identification 
and integration. 
Optimize your analysis first to get stable 
measuring conditions without a retention 
time drift. In any case it is better to have 
stable conditions, use this feature only as 
a last possibility. 
Calibration
If checked, the retention times will be up-
dated in the peak table after every calibra-
tion. 
All
If checked, the retention times will be up-
dated in the peak table after every determi-
nation. 
Recalculate only
Enable recalculation of chromatograms 
with values for 
Volume
,
Dilution
,
Amount
and
