Metrohm IC Net 2.3 User Manual
Page 238
6 Devices
IC Net 2.3
228
lution (D) in the sample queue:
Concentr. vol. = MV/D
Dilution
vol.
Dose Dilution vol. (DV) = MV *
(D-1)/D
Reagent
vol.
Dose
Reagent volume
(
R
) defined
on the tab Configuration/General.
-Reagent
vol.
Aspirate
Reagent volume
(
R
) defined
on the tab
Configuration
/
General
.
fill
Fills the dosino cylinder.
release
Prepares the dosino for changing
the dosing unit. The dosing cylin-
der is filled via the rinse port. The
stopcock is turned to the exchange
position (Port 2).
prepar.
Preparation cycle (PREP) for dosi-
nos. All tubing is rinsed and filled
completely.
empty
Tubing system and dosino cylinder
are emptied.
endVol
Dosino cylinder contents are
ejected.
compen.
Cancels the mechanical play be-
tween the dosing piston and spin-
dle.
port 1..4
Define new dosing port (Configura-
tion value Dos port is temporarily
overwritten).
Dosino
MSB1..3
rate
0.01…160 mL/min
Define new dosing rate for the cur-
rently used dosing port of the cor-
responding dosino (Configuration
value Rate is temporarily overwrit-
ten).
Tower Stir-
rer, Stirrer
MSB1..3
on
(infinite)
, off,
1...9999 s
Choose stirrer and select running
duration.
Membrane
Pump 1..2
on
(infinite)
, off,
1...999 s
Choose pump and select running
duration.