6 Devices

IC Net 2.3

228

lution (D) in the sample queue:

Concentr. vol. = MV/D

Dilution

vol.

Dose Dilution vol. (DV) = MV *

(D-1)/D

Reagent

vol.

Dose

Reagent volume

(

R

) defined

on the tab Configuration/General.

-Reagent

vol.

Aspirate

Reagent volume

(

R

) defined

on the tab

Configuration

/

General

.

fill

Fills the dosino cylinder.

release

Prepares the dosino for changing

the dosing unit. The dosing cylin-

der is filled via the rinse port. The

stopcock is turned to the exchange

position (Port 2).

prepar.

Preparation cycle (PREP) for dosi-

nos. All tubing is rinsed and filled

completely.

empty

Tubing system and dosino cylinder

are emptied.

endVol

Dosino cylinder contents are

ejected.

compen.

Cancels the mechanical play be-

tween the dosing piston and spin-

dle.

port 1..4

Define new dosing port (Configura-

tion value Dos port is temporarily

overwritten).

Dosino
MSB1..3
rate

0.01…160 mL/min

Define new dosing rate for the cur-

rently used dosing port of the cor-

responding dosino (Configuration

value Rate is temporarily overwrit-

ten).

Tower Stir-
rer, Stirrer
MSB1..3

on

(infinite)

, off,

1...9999 s

Choose stirrer and select running

duration.

Membrane
Pump 1..2

on

(infinite)

, off,

1...999 s

Choose pump and select running

duration.