6.28 844 UV/VIS Compact IC

IC Net 2.3

263

ample for the wavelength set for measuring channel

1:

akt

Int

zero

Int

Abs

1

1

log

λ

λ

=

where:

zero

Int

1

λ

The intensity measured when recording

the zero spectrum at the wavelength set

for measuring channel 1.

akt

Int

1

λ

The actual intensity measured at the

wavelength set for measuring channel 1.

If you are working with a reference channel (can be

activated on the subtab

Configuration

) the formula is

as follows:

refakt

Int

refzero

Int

akt

Int

zero

Int

Abs

λ

λ

λ

λ

log

log

1

1

=

where:

refzero

Int

λ

The intensity measured when recording

the zero spectrum at the wavelength set

for the reference channel.

refakt

Int

λ

The actual intensity measured at the

wavelength set for the reference channel.

Intensity

The intensity (y-axis) is given as a function of the wa-

velength (x-axis).

Zero

A click on

Zero

shows the last zero spectrum to have

been recorded (a zero spectrum is recorded by click-

ing on the

button on the subtab

Man-

ual/Detector

of the 844 control window).

The zero spectrum is a reference spectrum. It should

be recorded at a time at which there is no sample in

the measuring cell. It is used for calculating the ab-

sorption spectrum (see above formulas).

Wavelength, nm

Sets the shown x-axis range

Entry range:

190 ... 900 nm

, (standard settings 190

nm and 900 nm,

Upper limit

must be higher than

Lower limit

).

Spectrum

Sets the shown y-axis range

Standard settings:

0 ... 100.

Spline

Activates curve smoothing.

Send values to the 844 UV/VIS Compact IC.