4 processing – Metrohm IC Net 2.3 User Manual
Page 286

7 Methods
IC Net 2.3
276
bit smaller and wider after smoothing, but
their area does not change.
Raw data points
Number of data points per chromatogram
(read-only).
Smallest peak has ...
Number of data points per smallest peak
on the chromatogram (read-only). This in-
formation is valuable to perform the
Com-
press
procedure (data compression).
7.3.4 Processing
Tab
Processing
of
Method setup
window for definition of actions
that are performed during the run or when the chromatogram fin-
ishes. Automatic post-run data processing is especially useful for
chromatographic systems that include an autosampler.
Save chromatogram after the run
If checked, autosaving of the chroma-
togram will be performed on data acquisi-
tion ending.
while running save every ... min
Save the chromatogram during the run for
data safety at the time interval set (if
0
is
set, the chromatogram is not saved during
the run).
Update retention time on runs:
Update the retention times in the peak table
after the run. Tool to analyse difficult sam-
ples. Compensates a drift of the retention
time to enable correct peak identification
and integration.