7 Methods 
 
IC Net 2.3
276
bit smaller and wider after smoothing, but 
their area does not change. 
Raw data points
Number of data points per chromatogram 
(read-only). 
Smallest peak has ...
Number of data points per smallest peak
on the chromatogram (read-only). This in-
formation is valuable to perform the 
Com-
press
procedure (data compression).
7.3.4 Processing
Tab
Processing
of
Method setup
window for definition of actions
that are performed during the run or when the chromatogram fin-
ishes. Automatic post-run data processing is especially useful for 
chromatographic systems that include an autosampler. 
Save chromatogram after the run
If checked, autosaving of the chroma-
togram will be performed on data acquisi-
tion ending. 
while running save every ... min
Save the chromatogram during the run for 
data safety at the time interval set (if 
0
is
set, the chromatogram is not saved during 
the run). 
Update retention time on runs:
Update the retention times in the peak table 
after the run. Tool to analyse difficult sam-
ples. Compensates a drift of the retention 
time to enable correct peak identification 
and integration. 