6.21 837 IC Degasser
IC Net 2.3
171
Time program
On the
Program
tab of the
837 IC Degasser
window a user-defined
time program can be entered. This program is started automati-
cally as defined in the
Start mode
window (see 4.4.3) either at the
moment the determination is started (
Start with determination
) or at
the moment the sample is injected (
Start with inject
).
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see be-
low).
Enabled
Enable program start (a disabled program
is not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).