Integration events for peak detection – Metrohm IC Net 2.3 User Manual
Page 295
7.4 Integration
IC Net 2.3
285
If an integration event is added or modified, the
Edit integration
event
window is opened where the following parameters can be
set:
Time
Start time for integration event in min.
Event
List box with integration events to be se-
lected (see below).
Value
Parameter value for selected integration
event. This field appears for events that
demand input of an additional parameter.
Integration events for peak detection
Disable detection
Set this mode to stop detection of new
peaks. If a peak is available at the moment
of the event it is either finished (downslope
peaks) or rejected (upslope peaks).
Enable detection
Clear
Disable detection
mode.
Enable negative peaks
Enable detection of negative peaks. In
some cases this mode may result in insta-
bility of detection algorithm.
Disable negative peaks
Disable detection of negative peaks. This
event does not influence negative peaks
which already started.
Disable peak reject
Set this mode, if a small peak should not be
rejected because of its flat apex.
Enable peak reject
Clear
Disable peak reject
mode.