Metrohm IC Net 2.3 User Manual
Page 218

6 Devices
IC Net 2.3
208
Entry range:
0…999 µL
, default:
30
(with air
segment)
The minimum flush volume depends on
the volume of needle and needle tubing,
flush volumes of less than twice the vol-
ume of needle and tubing will result in
decreased performance.
Inj.
volume, µL
Programmable only for
Partial
and
Pick-up
modes, is equal to the
Loop volume
in
Flushed
mode. Is loaded from the sample
queue. The maximum injection volume de-
pends on the injection mode:
Partial:
50% of the loop volume
Pick-up:
(loop volume – 3
×
needle vol
ume)/2
Entry range:
1…1000 µL
Analysis time
Time between switching the injection valve
to inject and the start of processing the next
sample. Is loaded from the sample queue.
Entry range:
0:00:00…9:59:59 h:mm:ss
Delay time
Time
between
start determination
and start
of the
Triathlon methods/User program
, useful
e.g. for baseline equilibrium after a sup-
pressor step.
Entry range:
0:00:00…1:00:00 h:mm:ss
Apply parameter values selected under
Method values
to the
Injection method
.