6 Devices

IC Net 2.3

110

E1

Start potential for sweep.
Range:

-2 ... +2 V

.

E2

End potential for sweep.
Range:

-2 ... +2 V

.

Rate

Definition of scan rate in mV/s.
Selection:

1, 2, 5, 10, 20, 50 mV/s

Cycle mode

Definition of scan cycle:

full

double

scan

E1 ... E2 ... E1

half

single

scan

E1 ... E2

Selection:

full, half

Range

Selection of measuring range:

nA

10 ... 500 nA

.

µA

1 ... 200 µA

.

Offset

Offset of the zero point for the analog out-
put signal.
Selection:

-50 ... +50 %

(in steps of 10 %)

Data rate

Setting of the data acquisition rate.
Selection:

-1, 2, 5, 10 Hz

Actual

E1, E2

Display of current start and end potentials
in V.

Rate

Display of current scan rate in mV/s.

Cycle mode

Display of current scan cycle mode.

Range

Display of current measuring range.

Offset

Display of current offset.

Data rate

Display of current data rate.

Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (

*.smt

) as long as

the file is not saved.

Opens the 871 Scan shell window.