6 Devices
IC Net 2.3
80
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter entry
Meaning
Remote
0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter
1
,
0
or
*
. For
entry of the other values, move
the cursor in front of the value to
be changed and enter
1
,
0
or
*
.
Range
100, 200, 5000
µ
S/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05
µ
S/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on
Range
setting).
Zero on,
off
Switch on or off the auto-zero
function.
Polarity +,
–
Reset the polarity of the output
signal.
Mark
Trigger marking signal.
ValveA fill,
inject
Switch injection valve A to "
inject
"
or "
fill
" position.
ValveB fill,
inject
Switch injection valve B to "
inject
"
or "
fill
" position.
Suppressor
Switch suppressor module to the
next position.
StartDA
Start data acquisition.