6.10 709 IC Pump
IC Net 2.3
127
value during operation, it is switched off
immediately. At the same time, the
Pressure
field color in the 709 settings window and in
the
WATCH WINDOW
is changed to the
value set for
Out of range
.
The
set
minimum limit value should lie far
enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is
automatically switched off. At the same
time, the
Pressure
field color in the 709 set-
tings window and in the
WATCH WINDOW
is
changed to the value set for
Out of range
.
Send current parameters immediately to the
709 IC Pump (only available if pump is run-
ning). Parameters are not stored in the sys-
tem file (
*.smt
) as long as the file is not
saved.
Set current parameters and start 709 IC
Pump drive.
Stop 709 IC Pump drive.
Links
The
Links
tab of the
709 IC Pump
is used for COM port selection
and settings (details see section 5.2.4 Links).
Advanced
The
Advanced
tab of the
709 IC Pump
window is only available for a
connected system.