beautypg.com

Events configuration – Metrohm IC Net 2.3 User Manual

Page 58

background image

5 Interfaces

IC Net 2.3

48

Program END,

RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the

chromatogram duration. Additional

steps after this flag are not allowed.

The

RESET

flag is used to reset the

parameters to the system startup

values.



Events configuration

On the

Events configuration

subtab user-defined event commands

can be defined, which can be inserted into a time program after
being activated with

.

Name (1

st

column)

User-definable name of the event com-
mand (e.g.

Start_766

).

Event command (2

nd

column)

Setting the event output lines 1

7.

Selection:

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(pulse, pulse length 150 ms)

*

(leave line in current status)

For entry of the first value, enter

1

,

0

,

p

or

*

.

For entry of the other values, move the cur-
sor in front of the value to be changed and
enter

1

,

0

,

p

or

*

.

Add new remote command.

Delete selected remote command.

Activate the defined remote commands for
insertion into the time program.