beautypg.com

Time program – Metrohm IC Net 2.3 User Manual

Page 177

background image

6.20 833 IC Liquid Handling Unit

IC Net 2.3

167

Suppressor

(available only for

833 IC Suppressor Unit

)

Rotate suppressor to the next position. The
time since the last step action is displayed
in the field beside the

button.

Actuator

(available only for

833 IC Ultra Filtration Unit

)

Rotate sample preparation module to the
next position. The time since the last step
action is displayed in the field beside the

button.

Start pump with startup hardware

Automatic start of pump drive with

Startup

hardware

.


Time program

On the

Program

tab of the

833 IC Liquid Handling Unit

window a

user-defined time program can be entered. This program is
started automatically as defined in the

Start mode

window (see

section 4.4.3) either at the moment the determination is started
(

Start with determination

) or at the moment the sample is injected

(

Start with inject

).

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min