Method parameter – Metrohm IC Net 2.3 User Manual
Page 87

6.4 819 IC Detector
IC Net 2.3
77
Damping
Display of electronic damping of the analog 
output. 
Remote lines
Display value of remote line settings 1
…
8.
Zero 
Triggers the auto-zero function.
Zero 
Switch off the auto-zero function.
 
 
Method parameter
The
Method parameter
tab of the
819 IC Detector
window contains
the method startup values for the instrument. These values are 
sent and applied to the 819 IC Detector by 
Start determination
and
Startup hardware
, or if the values are sent manually with
.
Range
Measuring range in
µ
S/cm
Selection:
100, 200, 500
µ
S/cm
and
1, 2, 5, 10
mS/cm
Full
scale
The full scale range (operating range) sets 
the desired sensitivity for the analog output. 
The possible values of the full scale range 
in 
µ
S/cm depend on the preset measuring
Range
.
Ion
Combination of polarity and temperature 
coefficient: 
Anion
polarity = +, temperature coeff. = 2.5
Cation
polarity = –, temperature coeff. = 1.5