beautypg.com

Method parameter – Metrohm IC Net 2.3 User Manual

Page 87

background image

6.4 819 IC Detector

IC Net 2.3

77

Damping

Display of electronic damping of the analog
output.

Remote lines

Display value of remote line settings 1

8.

Zero

Triggers the auto-zero function.

Zero

Switch off the auto-zero function.


Method parameter

The

Method parameter

tab of the

819 IC Detector

window contains

the method startup values for the instrument. These values are
sent and applied to the 819 IC Detector by

Start determination

and

Startup hardware

, or if the values are sent manually with

.

Range

Measuring range in

µ

S/cm

Selection:

100, 200, 500

µ

S/cm

and

1, 2, 5, 10

mS/cm

Full

scale

The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
in

µ

S/cm depend on the preset measuring

Range

.

Ion

Combination of polarity and temperature
coefficient:

Anion

polarity = +, temperature coeff. = 2.5

Cation

polarity = –, temperature coeff. = 1.5