Method parameter – Metrohm IC Net 2.3 User Manual
Page 87
6.4 819 IC Detector
IC Net 2.3
77
Damping
Display of electronic damping of the analog
output.
Remote lines
Display value of remote line settings 1
…
8.
Zero
Triggers the auto-zero function.
Zero
Switch off the auto-zero function.
Method parameter
The
Method parameter
tab of the
819 IC Detector
window contains
the method startup values for the instrument. These values are
sent and applied to the 819 IC Detector by
Start determination
and
Startup hardware
, or if the values are sent manually with
.
Range
Measuring range in
µ
S/cm
Selection:
100, 200, 500
µ
S/cm
and
1, 2, 5, 10
mS/cm
Full
scale
The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
in
µ
S/cm depend on the preset measuring
Range
.
Ion
Combination of polarity and temperature
coefficient:
Anion
polarity = +, temperature coeff. = 2.5
Cation
polarity = –, temperature coeff. = 1.5