Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual
Page 90
![background image](/manuals/401804/90/background.png)
Boonton 4500B RF Peak Power Analyzer
Operation
4-18
Table 4-3. Chan # > Calibration > Submenu (continued)
Menu Item
Function
(Type)
Selections
(SCPI cmd)
Initiate Zeroing by pressing the Chan # > Calibration > Zeroing menu
key. During the Zeroing process, the message line will read ―Zeroing
Channel #.‖ When complete, the message line reads ―Zeroing
Completed Successfully.‖
Reactivate any external signal applied to the sensor.
CALibration[1|2]:ZERO
Note
If the SRQ indicator appears at the end of the AutoCal, Fixed Cal or Zeroing procedures, refer to
Subsection 5.4 SRQ Operation. If an error message appears on the display during the Autocal, Fixed Cal
or Zeroing procedures, refer to Appendix A Error Messages.