beautypg.com

Table 4-9c. trig > menu (statistical), Table 4-9c. trig > menu (statistical) -50, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 122: Operation

background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-50

Table 4-9C. Trig > Menu (statistical)

Menu Item

(Type)

Selections

Function


Gating

Stat

Select free running or gated statistical mode

(Toggle)

Free Run, Pulse Gated


Free Run – use this mode to build a
distribution function that contains all of the
samples collected between the Start and
Stop measurement commands.

Pulse Gated – use this mode to build a
distribution function that contains only the
samples within the time interval defined by
Gate 1 and Gate 2 that are collected between
the Start and Stop measurement commands.
This mode is used to measure the statistics
of a burst that can be synchronized with the
trigger and timebase system.


TRIGger:STATistical:GATIng {ON, OFF}

OFF = Free Run; ON = Pulse Gated



Trig Setup

Gated Stat

Setup the pulse gate trigger parameters

(Submenu)

MENU


This menu is identical to Trig > menu above,
Table 4-9 and its submenu, Table 4-9A.
Trig > B Trigger. The paths will appear as
Trig > Trig A and Trig > Trig A > B Trig.
Stat*
indicates a parameter that applies to
this setup.



Time Setup

Gated Stat

Setup the pulse gate timebase parameters

(Submenu)

MENU

This menu is identical to the Time > menu
above, Table 4-8A. The path will appear as Fig. 4-27f. Trig > menu
the Trig > Time Settings > menu. Stat*

Statistical Mode

indicates a parameter that applies to this
setup.


Gate 1

Gated Stat

Mark the beginning of the statistical time gate

(Numeric)

Continuous range:

With the burst of interest synchronized in the Pulse Preview window
position the Gate 1 vertical cursor at the start of the region to be
included in the distribution function.


TRIGger:STATistical:GATE[1] (sec)