System:communicate:gpib:listen, System:communicate:gpib:talk, System:communicate:logging – Boonton 4500B Peak Power Meter User Manual
Page 281: System:communicate:serial:baud, System:communicate:gpib:listen -81, System:communicate:gpib:talk -81, System:communicate:logging -81, System:communicate:serial:baud -81, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/281/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-81
SYSTem:COMMunicate:GPIB:LISTen
Description:
Set or return the 4500B as LISTENER line termination parameter. EOIONLY is the default and
the preferred setting. The 4500B always responds to EOI when listening. When EOI is not in use
or not available from the bus controller, select the CR or LF character for the end of string (EOS)
indicator and set the controller to send the same character as the EOS. If this is not done, GPIB
communication will not be possible.
Syntax:
SYSTem:COMMunicate:GPIB:LISTen
Argument:
Valid Modes:
Any
SYSTem:COMMunicate:GPIB:TALK
Description:
Set or return the 4500B as TALKER line termination parameter. EOIONLY is the default and the
preferred setting. The 4500B always sends EOI when talking. When the bus controller does not or
cannot respond to EOI, select the CR, LF or CRLF character(s) for the end of string (EOS)
indicator and set the controller to accept the same character(s) as the EOS. If this is not done,
GPIB communication will not be possible.
Syntax:
SYSTem:COMMunicate:GPIB:TALK
Argument:
Valid Modes:
Any
SYSTem:COMMunicate:LOGGing
Description:
Set or return the status of GPIB command logging. Logging saves to a file all GPIB messages
received and sent for review.
Syntax:
SYSTem:COMMunicate:LOGGing
Argument:
Valid Modes:
Any
SYSTem:COMMunicate:SERial:BAUD
Description:
Set or return the Baud rate of the 4500B RS232 port (COM1).
Syntax:
SYSTem:COMMunicate:SERial:BAUD
Argument:
Valid Modes:
Any