Display:pctofst, Display:pctscale, Display:persistence – Boonton 4500B Peak Power Meter User Manual
Page 227: Display:plabel:mode, Display:pctofst -27, Display:pctscale -27, Display:persistence -27, Display:plabel:mode -27, Boonton 4500b rf peak power analyzer, Remote operation
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-27
DISPlay:PCTOFST
Description:
Set or return the horizontal trace percent offset of the statistical linear display mode.
Syntax:
DISPlay:PCTOFST
Argument:
Valid Modes:
Statistical mode only
Restrictions:
The sum of (PCTSCALE * 10 div) and PCTOFST must not be greater than 100%.
DISPlay:PCTSCALE
Description:
Set or return the horizontal trace percent/div scale in the statistical linear display mode. Range
settings are fixed in a 1, 2, 5 sequence, and if the argument does not match one of the settings, it
will be forced to the next highest entry.
Syntax:
DISPlay:PCTSCALE
Argument:
Valid Modes:
Statistical mode only
Restrictions:
The sum of 10*PCTSCALE and PCTOFST must not be greater than 100%.
DISPlay:PERSistence
Description:
Set or return the status of the display persistence function. Affects CH1, CH2, TRIG1, TRIG2 and
CHM. The ON argument sets the persistence mode and forces the dot trace type. See
DISPlay:TRACe:TYPE.
Syntax:
DISPlay:PERSistence
Argument:
Valid Modes:
All
DISPlay:PLABel:MODE
Description:
Set or return the status of the plot label display function. The plot label consists of 4 user
programmable text lines in the trace box.
Syntax:
DISPlay:PLABel:MODE
Argument:
Valid Modes:
All